Unit of measurement | ml/100g [@ 20°C and 1 atm]
(milliliter of hydrogen per 100 grams of liquid aluminium ) |
Sensor type | Electrochemical. Solid-state reference |
Reproducibility | ±0.01 ml/100g or 3% of hydrogen concentration |
Response time | Initial result: 5 minutes typically. Subsequent response: < 5 seconds |
Temperature of operation | 620°C to 800°C |
Analyser | 400 (W) x 400 (H) x 200 (D) mm. / 16 kg. |
Probe | Standard length: 800mm.Maximum depth of immersion: 300mm Diameter: 6mm |
Probe cable (std) | 3000mm |
Refillable gas bottle (300ml) | Dry argon or nitrogen. Max fill pressure: 120 bar |
Thermocouple | Type K |
Specified limits of error | 1.1 deg C or 0.4% - whichever the greater |
Power supply (external unit) | AC 85 - 265 V, 47 - 63 Hz |
Typical power consumption | < 10 W (probe heater off, battery discharging) |
Max. power consumption | 210 W (probe heater on, battery charging) |
Battery chemistry | Lithium ion |
Battery life | 8 - 14 hours (depending on screen brightness) |
Screen lighting | LED |
Screen resolution | 800 x 480 WVGA, TFT colour LCD |
Screen size | 7” (177mm) diagonal |
Touch screen type | Analogue-resistive with 4H hardness |
Ethernet interface | 10/100 Base-T, RJ45 |
Communication protocol | MODBUS TCP |
USB | 2.0 full-speed (Type A) |
Operating Temperature | 0 - 60 deg C |
Humidity | 5 - 95% non-condensing |