EMC

Hycal 1000 - Technical

ParameterSpecification
Unit of measurementml/100g [@ 20°C and 1 atm] (milliliter of hydrogen per 100 grams of liquid aluminium )
Sensor typeElectrochemical. Solid-state reference
Reproducibility±0.01 ml/100g or 3% of hydrogen concentration
Response timeInitial result: 5 minutes typically. Subsequent response: < 5 seconds
Temperature of operation620°C to 800°C
Analyser400 (W) x 400 (H) x 200 (D) mm. / 16 kg.
ProbeStandard length: 800mm.Maximum depth of immersion: 300mm Diameter: 6mm
Probe cable (std)3000mm
Refillable gas bottle (300ml)Dry argon or nitrogen. Max fill pressure: 120 bar
ThermocoupleType K
Specified limits of error1.1 deg C or 0.4% - whichever the greater
Power supply (external unit)AC 85 - 265 V, 47 - 63 Hz
Typical power consumption< 10 W (probe heater off, battery discharging)
Max. power consumption210 W (probe heater on, battery charging)
Battery chemistryLithium ion
Battery life8 - 14 hours (depending on screen brightness)
Screen lightingLED
Screen resolution800 x 480 WVGA, TFT colour LCD
Screen size7” (177mm) diagonal
Touch screen typeAnalogue-resistive with 4H hardness
Ethernet interface10/100 Base-T, RJ45
Communication protocolMODBUS TCP
USB2.0 full-speed (Type A)
Operating Temperature0 - 60 deg C
Humidity5 - 95% non-condensing