Hycal Mini Mk3
Technical specifications
| Parameter | Specification |
|---|---|
| Unit of measurement | ml/100g [@ 20°C and 1 atm] (milliliter of hydrogen per 100 grams of liquid aluminium ) |
| Sensor type | Electrochemical. Solid-state reference |
| Reproducibility | ±0.01 ml/100g |
| Response time | Initial result: 5 minutes typically. Subsequent reading: every 2 seconds |
| Temperature of operation | 650°C to 800°C |
| Analyser | 360mm x 300mm x 170mm / 10.5 kg |
| Probe | Standard length: 800mm. Maximum depth of immersion: 300mm Diameter: 6mm |
| Probe cable (std) | 3000mm |
| Refillable gas bottle (300ml) | N/A - Probe Adaptor required |
| Thermocouple | Type K |
| Power supply (external unit) | AC 85 - 265 V, 47 - 63 Hz |
| Typical power consumption | < 8W |
| Max. power consumption | 210W |
| Battery chemistry | Lithium polymer |
| Battery life | 11 hours (depending on screen brightness) |
| Screen lighting | LED (white on black background) |
| Screen resolution | 800 x 480 TFT colour LCD |
| Screen size | 5" (125mm) diagonal |
| Screen type | LCD graphical display |
| Operating Temperature | 0 - 50 deg C |
| Humidity | 5-95% non-condensing |